Data taken with ARC controller (April 2007), and with the array in upside-down (note triangle feature on array is now at the bottom!).
1 second dark : A standard short array-test dark. Should see “zero counts”. .sdf file available here
100 second dark : A standard long array-test dark. Should see 15-20 counts. .sdf file available here
4 second HK flat, 4 pix slit : A standard HK spectroscopy flat. Should see counts in range 1000-8000. .sdf file available here
15 second HK arc, 4 pix slit : A standard HK spectroscopy arc. Should see strong, narrow (4-pixel wide) lines. .sdf file available here
40 second short-K IFU flat : A standard short-K IFU flat. Should see counts in range 1000-8000. .sdf file available here
30 second short-K IFU arc : A standard short-K IFU arc. Should see strong, narrow (2-pixel wide) lines. .sdf file available here
Coronagraphic mask : K-band Coron exposure. Taken during day (night sky should be smoother). .sdf file available here
Coronagraphic mask with IRPOL : K-band Coron + Pol exposure. Taken during day (night sky should be smoother). .sdf file available here
THERMAL IMAGING
SATURATING : High counts (25,000-30,000) top and bottom, with low or maybe even negative counts in the middle.
NOT SATURATING : Counts should be 10,000-25,000
WHAT YOU SHOULD NOT SEE
Disabled array data taken with ARC controller; all other images taken with old EDICT system, although the images should look the same if these problems re-occur.
Thermal gradient : A bad 100 sec dark showing the gradient. More details given here
Ice on the UIST window : A daytime K-band image of inside of dark dome Details on clearing ice given here
Image Rotator at 0 degrees : For imaging should be at -90 degrees! Above can see vignetted corners, because the image rotator actually rotates by 45 degrees for every 90 (uses two folding mirrors).
Array disabled : All zeros – Have forgotten to enable the array
Bad channel : Low-frequency sine wave along every 8th column in two quadrants (from summer 2007 – SDF frame )… …
ENGINEERING (UCLIENT DATA)
Data taken with ARC controller (Dec 2006), though while still fixing boards, etc. (so striping in bottom-right quadrant not normal).
Stare frame Full array (UCLIENT default settings with srr1024)SDSU-stare.fits
CDS frame Full array (UCLIENT default settings with cds1024)SDSU-cds.fits
NDR frame Full array (UCLIENT default settings with ndr1024)SDSU-ndr.fits